Origin / Brand: American LGR company
Model: IWA-35-EP
Operational principle:
Off axis integrating cavity continuous spectral output technique (OA-ICOS)
Performance index
Sample volume: <10 µ L
Number of samples / day: 90 samples / day
Specular reflectance: better than 99.999%
Salinity: < 4%
Repeatability / accuracy (1-sigma):18o/16o:0.2 ‰; D/h:0.6 ‰ (30 samples / day)
Output: digital (RS232), Ethernet
Data storage: built in hard disk (20g)
Sampling temperature: 0-50 ° C
Operating temperature: 5-45 ° C
Preheating time: 2 minutes
Inlet / outlet connection: quick replaceable injection diaphragm, 1/4 "connector sleeve for dry gas cleaning
Power demand: 115/230 VAC; 50/60 Hz; 80 W (analyzer only); 115/230 VAC; 50/60 Hz; 120 W (with external pump)
Size: 10 "H × 38 "W × 14 "d
Main functions:
It can be used for the analysis and monitoring of hydrogen and oxygen isotopes in rainwater, rivers, groundwater and other liquid samples
Application fields:
Research and development in the field of earth environment and biology, water isotope monitoring of industrial wastewater, surface water, groundwater and soil water and its environmental significance
Research projects that can be carried out (test projects, problems solved):
1. analysis and detection of hydrogen and oxygen isotopes in groundwater, surface water, drinking water, sewage and other water bodies.
2. isotope analysis of liquid water in sediments;
3. analysis and detection of hydrogen and oxygen isotopes in liquid water of organisms and plants.